CIC-D100 ion chromatograph is a classic product of SHINE, which has been accepted by many customers.Based on the latest requirements of users, a newly upgraded CIC-D100 for conventional detection came into being.The new IC not only can detect anions, cations and other polar substances in different matrix samples, but also separate ions with 4 orders of magnitude difference.Compared with the previous one, it is more accurate and reliable. One key switch and intelligent maintenance functions are added to give users a better experience. It is suitable for third-party testing institutions, enterprises, environmental protection, chemical industry, mining and metallurgy and other fields.
Pump | |
---|---|
Maximum Pressure | 42 MPa (Stainless steel) |
Flow Rate Range | 0.001 ~ 9.999 mL/min |
Injector | |
Maximum Pressure | 35 MPa (PEEK) |
External Diameter of The Tube | 1/16" |
Column Heater | |
Operating Temperature Range | 20 ~ 60℃ (68 ~ 140℉) |
Temperature Accuracy | ± 1 ℃ |
Temperature Stability | ≤ 0.05 ℃/h |
Conductivity Detector | |
Cell Volume | ≤ 0.8 μL |
Detection Range | 0 ~ 35000 μS/cm |
Baseline Noise | ≤ 0.001 μS/cm |
Baseline Drift | ≤ 0.02 μS |
Power Requirements | 150 W |
Dimensions (L x W x H) | 310 x 430 x 530 mm |
Net Weight | 22 Kg |
Gross Weight | 29 Kg |