Technical specifications
|
|
---|---|
Flow rate
|
500 Nml/min ± 10% a 2000 m asl
|
Technology
|
PSA CLAIND
|
Output pressure
|
0 - 6 bar/0 - 87 psi ± 3%
|
Purity
|
> 99,9995 %
|
Moisture
|
< -50 °C dew-point ATP ± 2°C ATP
|
Air input characteristics
|
Particulate ≤ 0,01 mm
Oil vapors ≤ 0,01 mg/m3
Moisture ≤ 3° C dew point ATP
min pressure: 8 bar
Air flow rate: 6 Nl/min
|
Electrical supply x nigen gc
|
115/230 Vac ±10% 60-50 Hz 50 w
|
Electrical supply x nigen gc hc
|
115 Vac ±10% 60 Hz 200 W
230 Vac ±10% 50 Hz 200 W
|
Index of protection
|
IP20
|
Noise
|
< 50 dBA for the 98% of operation time
< 60 dBA for 2 sec. each cycle
|
Operating temperature
|
5°C - 40°C/40°F - 100°F
|
Environmental humidity
|
max 90% without condensation
|
Max altitude
|
2000 m asl
|
Dimensions
|
Height: 40 cm + 7 cm/15.8” + 2.8”
Width: 38 cm/14.9”
Depth: 53 cm/20.8”
|
Weight
|
GC version: 28 kg/62 lbs GC HC version: 32 kg/70.5 lbs
|
Pneumatic connections
|
Compressed air inlet: G 1/8” female
Nitrogen outlet: G 1/8” female
|
Technical specifications
|
|
---|---|
Flow rate
|
500 Nml/min ± 10% a 2000 m asl
|
Technology
|
PSA CLAIND
|
Output pressure
|
0 - 6 bar/0 - 87 psi ± 3%
|
Purity
|
> 99,9995 %
|
Moisture
|
< -50 °C dew-point ATP ± 2°C ATP
|
Air input characteristics
|
Particulate ≤ 0,01 mm
Oil vapors ≤ 0,01 mg/m3
Moisture ≤ 3° C dew point ATP
min pressure: 8 bar
Air flow rate: 6 Nl/min
|
Electrical supply x nigen gc
|
115/230 Vac ±10% 60-50 Hz 50 w
|
Electrical supply x nigen gc hc
|
115 Vac ±10% 60 Hz 200 W
230 Vac ±10% 50 Hz 200 W
|
Index of protection
|
IP20
|
Noise
|
< 50 dBA for the 98% of operation time
< 60 dBA for 2 sec. each cycle
|
Operating temperature
|
5°C - 40°C/40°F - 100°F
|
Environmental humidity
|
max 90% without condensation
|
Max altitude
|
2000 m asl
|
Dimensions
|
Height: 40 cm + 7 cm/15.8” + 2.8”
Width: 38 cm/14.9”
Depth: 53 cm/20.8”
|
Weight
|
GC version: 28 kg/62 lbs GC HC version: 32 kg/70.5 lbs
|
Pneumatic connections
|
Compressed air inlet: G 1/8” female
Nitrogen outlet: G 1/8” female
|
Technical specifications
|
|
---|---|
Flow rate
|
6 Nml/min ± 10% a 2000 m asl
|
Technologies employ
|
PSA CLAIND
|
Output pressure
|
0 - 7 bar/0 - 100 psi ± 3%
|
Purity
|
> 99,9995 %
|
Oil free compressor
|
Included
|
Electrical supply
|
115 Vac ±10% 60 Hz 200 W
230 Vac ±10% 50 Hz 200 W
|
Index of protection
|
IP20
|
Noise
|
< 50 dBA during the stand by of the compressor
|
Operating temperature
|
5°C fino 35°C (41°F fino 95°F)
|
Environmental humidity
|
max 90% without condensation
|
Dimensions
|
Height: 132 cm/52”
Width: 44 cm/17.3”
Depth: 110 cm/43.3”
|
Weight
|
220 kg/489 lbs
|
Pneumatic connections
|
Compressed air inlet: G 1/4” female
Nitrogen outlet: G 1/4” female
|
Established in 1979, Claind produces hydrogen, nitrogen and zero air generators. Its headquarters is in Lenno, on the shores of Lake Como, where all product development, production and administration activities are carried out. The customers who Claind targets can be broken down into three main categories: analysis laboratories, food and beverage companies and laser cutting companies.
Over the years Claind's staff has received and paid back the confidence of more than 3,000 customers with over 20,000 installations worldwide.
The principle that guides every business decision is the desire to achieve customer satisfaction, in this way strengthening a relationship based on trust that goes far beyond mere contacts for technical and commercial aspects.